Ultrasonic scaler' troubleshooting and notes
Fault | Possible cause | Solutions |
The scaling tip doesn't vibrate and there is no water flowing out when stepping on the foot switch. | The power plug is in loose contact. | Make the plug insert to the socket well. |
The foot switch is in loose contact. | Insert the foot switch to its socket tightly. | |
The fuse in the main unit is broken. | Contact our dealers or us. | |
The scaling tip doesn't vibrate but there is water flowing out when stepping on the switch. | The tip is in loose contact. | Screw the tip on the handpiece tightly. |
The connect plug between the handpiece and the circuit board is in loose contact. | Contact our dealers or us. | |
Something wrong with the handpiece. | Send the handpiece to our company to repair. | |
Something wrong with the cable. | Contact our dealers or us. | |
The scaling tip vibrates but there is no spray when stepping on the foot switch. | The water control knob is not on. | Turn on the water contrl knob. |
The vibration of the tip becomes weak. | The tip hasn't been screwed on to the handpiece tightly. | Screw the tip on the handpiece tightly. |
The tip is loose because of vibration. | Screw on the tip tightly. | |
The coupling between the handpiece and the cable isn't dry. | Dry it by the hot air. | |
There is water seeping from the coupling between the handpiece and cable. | The waterproof "O"ring is damaged. | Change a new waterproof "O"ring. |
The u-file doesn't vibrate. | The screw is loose. | Tighten it. |
Endochuck is damaged. | Change a new one. | |
There is noise coming from the endochuck. | The screw is loose. | Tighten it. |
Peristaltic pump leaks. | The inner water pipe cracks. | Replace a new peristaltic pump. |
LED light don't work | Poor contact. | Contact tightly. |
Something wrong with LED light. | Change a new one. | |
LED lamp installed Backwards. | Please install the "+" of the LED lamp to the "+" of the handpiece. | |
There is no water coming out from the handpiece (automatic water supply mode). | There is air in the water pipe. | Turn the water control to the Max, reinsert the bottle. |